6200T Total Sulfur Analyzer

Maintenance

Teledyen Analytical Instruments

235

9. Wait 5 minutes.

If the gauge pressure drops >1 psi within 5 minutes, then the hydrocarbon scrubber has
an internal leak and must be replaced. Contact Teledyne Analytical Instruments’s
Customer Service.