Setup Menu

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

126

To make these manual adjustments, the AOUT auto-calibration feature must be turned
OFF (refer to Section 5.9.3.2). Activate the ANALOG I/O
CONFIGURATION
MENU from the DIAG Menu (refer to Figure 5-16), then press:


DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AOUTS CALIBRATED:

NO

< SET

SET>

CAL

EXIT

DIAG AIO

CONC_OUT_1 :5V, NO CAL

< SET SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_1 CALIBRATED: YES

< SET

CAL

EXIT

DIAG AIO

CONC_OUT_1 OVERRANGE: ON

SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_1 CALIBRATED: NO

< SET

CAL

EXIT

DIAG AIO

CONC_OUT_1 VOLT–Z : 0 mV

U100 UP10 UP DOWN DN10 D100

ENTR EXIT

These control buttons increase / decrease the

analog output by 100, 10 or 1 counts.

Continue adjustments until the voltage measured

at the output of the analyzer and/or the input of

the recording device matches the value in the

upper right hand corner of the display to the

tolerance listed in Table 6-10.

The concentration display will not change. Only

the voltage reading of your voltmeter will change.

DIAG AIO

CONC_OUT_1 VOLT–S : 4500 mV

U100 UP10 UP DOWN DN10 D100

ENTR EXIT

DIAG AIO

CONC_OUT_1 REC OFS: 0 mV

< SET

SET>

EDIT

EXIT

If AutoCal is ON, go to

Section 6.7.3

EXIT

ignores the

new setting.

ENTR

accepts the

new setting.

DIAG AIO

CONC_OUT_1 AUTO CAL: OFF

< SET

SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_1 RANGE: 5V

SET>

EDIT

EXIT

Press SET> to select the analog output channel to be configured:

DISPLAYED

AS

=

CHANNEL

CONC_OUT_1

=

A1

CONC_OUT_2

=

A2

TEST

OUTPUT

=

A4

Figure 5-24:

Analog Output – Voltage Adjustment