Analog inputs (xin1…xin8) option configuration – Teledyne 6200T - Sulfides Analyzer User Manual
Page 130
Setup Menu
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
130
5.9.3.6. AIN Calibration
This is the sub-menu to conduct the analog input calibration. This calibration should
only be necessary after major repair such as a replacement of CPU, motherboard or
power supplies. Navigate to the ANALOG I/O CONFIGURATION MENU from the
DIAG Menu (refer to Figure 5-16), then press:
Exit at any time to
return to the main
DIAG
menu
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
Continue pressing SET> until …
DIAG AIO
AIN CALIBRATED:
NO
< SET
SET>
CAL
EXIT
DIAG AIO
AIN CALIBRATED:
YES
< SET
SET> CAL
EXIT
Instrument
calibrates
automatically
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
DIAG AIO
CALIBRATING A/D ZERO
CALIBRATING A/D SPAN
Figure 5-28:
DIAG – Analog Output – AIN Calibration
5.9.3.7. Analog Inputs (XIN1…XIN8) Option Configuration
To configure the analyzer’ optional analog inputs define for each channel:
gain (number of units represented by 1 volt)
offset (volts)
engineering units to be represented in volts (each press of the touchscreen
button scrolls the list of alphanumeric characters from A-Z and 0-9)
whether to display the channel in the Test functions