Analog inputs (xin1…xin8) option configuration – Teledyne 6200T - Sulfides Analyzer User Manual

Page 130

Advertising
background image

Setup Menu

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

130

5.9.3.6. AIN Calibration

This is the sub-menu to conduct the analog input calibration. This calibration should
only be necessary after major repair such as a replacement of CPU, motherboard or
power supplies. Navigate to the ANALOG I/O CONFIGURATION MENU from the
DIAG Menu (refer to Figure 5-16), then press:

Exit at any time to

return to the main

DIAG

menu

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

Continue pressing SET> until …

DIAG AIO

AIN CALIBRATED:

NO


< SET

SET>

CAL

EXIT

DIAG AIO

AIN CALIBRATED:

YES


< SET

SET> CAL

EXIT

Instrument

calibrates

automatically

Exit to return to the

ANALOG I/O

CONFIGURATION

MENU

DIAG AIO

CALIBRATING A/D ZERO


CALIBRATING A/D SPAN

Figure 5-28:

DIAG – Analog Output – AIN Calibration

5.9.3.7. Analog Inputs (XIN1…XIN8) Option Configuration

To configure the analyzer’ optional analog inputs define for each channel:

 gain (number of units represented by 1 volt)
 offset (volts)
 engineering units to be represented in volts (each press of the touchscreen

button scrolls the list of alphanumeric characters from A-Z and 0-9)

 whether to display the channel in the Test functions