Inet (ethernet), Com1 and com2 (mode, baud rate and test port) – Teledyne 6200T - Sulfides Analyzer User Manual
Page 112
Setup Menu
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
112
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
COM2
EXIT
SETUP X.
MACHINE ID: 100 ID
0 1 0 0
ENTR EXIT
Toggle to cycle
through the available
character set: 0-9
ENTR accepts the new
settings
EXIT ignores the new
settings
Figure 5-14:
COMM – Machine ID
The ID can be any 4-digit number and can also be used to identify analyzers in any
number of ways (e.g. location numbers, company asset number, etc.)
5.7.2. INET (Ethernet)
Use SETUP>COMM>INET to configure Ethernet communications, whether manually
or via DHCP. Please see Section 6.5 for configuration details.
5.7.3. COM1 and COM2 (Mode, Baud Rate and Test Port)
Use the SETUP>COMM>COM1[COM2] menus to:
Configure communication modes (Section 6.2.1)
View/set the baud rate (Section 6.2.2)
Test the connections of the com ports (Section 6.2.3).
Configuring COM1 or COM2 requires setting the DCE DTE switch on the rear panel.
Section 6.1 provides DCE DTE information.