Setup Menu
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
108
If the password feature is enabled, then when entering either Calibration or Setup Mode,
the default password displayed will be 000, and the new password must be input.
Example follows for Calibration Mode:
SAMPLE
ENTER SETUP PASS : 0
0 0 0
ENTR
EXIT
SAMPLE RANGE = 500.0 PPB SO2 =XXX.X
< TST TST >
CAL CALZ CALS
SETUP
M-P CAL RANGE = 500.0 PPB SO2 =XXX.X
< TST TST > ZERO
CONC
EXIT
Prompts
password
number
SAMPLE
ENTER SETUP PASS : 0
1 0 1
ENTR
EXIT
Press
individual
buttons to set
101
Continue calibration process …
Figure 5-10:
SETUP – Enter Calibration Mode Using Password