Control inputs – Teledyne 6200T - Sulfides Analyzer User Manual

Page 178

Advertising
background image

Remote Operation

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

178

Table 8-1: Status Output Pin Assignments

CONNECTOR PIN

STATUS

CONDITION (ON=CONDUCTING)

1

System Ok

ON if no faults are present.

2

Conc Valid

ON if concentration measurement is valid, OFF when invalid.

3

High Range

ON if unit is in high range of any AUTO range mode.

4

Zero Cal

ON whenever the instrument is in ZERO calibration mode.

5

Span Cal

ON whenever the instrument is in SPAN calibration mode.

6

Diag Mode

ON whenever the instrument is in DIAGNOSTIC mode.

7-8

Unused

D

Emitter Bus

The emitters of the transistors on pins 1-8 are bussed together. For most
applications, this pin should be connected to the circuit ground of the
receiving device.

+

Dc Power

+ 5 VDC source, 30 mA maximum (combined rating with Control Inputs)

Digital Ground

The ground from the analyzer’s internal, 5/±15 VDC power supply.

8.1.2. Control Inputs

Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes
are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear
panel. These are opto-isolated, digital inputs that are activated when a 5 VDC signal
from the “U” pin is connected to the respective input pin.

Table 8-2: Control Input Pin Assignments

INPUT STATUS

CONDITION

WHEN

ENABLED

A External

Zero

Cal

Zero calibration mode is activated. The mode field of the display will read
ZERO CAL R.

B External

Span

Cal

Span calibration mode is activated. The mode field of the display will read
SPAN CAL R.

C

Unused

D

Unused

E

Unused

F

Unused

Digital Ground

Provided to ground an external device (e.g., recorder).

U

DC Power For Input

Pull Ups

Input for +5 VDC required to activate inputs A - F. This voltage can be taken
from an external source or from the “+” pin.

+

Internal +5v Supply

Internal source of +5V which can be used to activate inputs when connected
to pin U.

There are two methods to activate control inputs. The internal +5V available from the
“+” pin is the most convenient method (Figure 8-2). However, to ensure that these
inputs are truly isolated, a separate, external 5 VDC power supply should be used
(Figure 8-3).

Advertising