5.9.3.4. Analog Output Offset Adjustment

Some analog signal recorders require that the zero signal to be significantly different
from the baseline of the recorder in order to record slightly negative readings from noise
around the zero point. This can be achieved in the 6200T by defining a zero offset, a
small voltage (e.g., 10% of span), which can be added to the signal of individual output
channels by activating the ANALOG I/O CONFIGURATION MENU from the DIAG
Menu (refer to Figure 5-16), then press:

DIAG AIO

RECORD OFFSET: 0 MV

+ 0 0 0 0

ENTR EXIT

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AOUTS CALIBRATED:

NO

< SET

SET>

CAL

EXIT

DIAG AIO

CONC_OUT_2 RANGE: 5V

SET>

EDIT

EXIT

Set the recorder

offset (in mV) of

the selected

channel

DIAG AIO

CONC_OUT_2:5V, CAL

< SET SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_2 REC OFS: 0 mV

< SET

SET> EDIT

EXIT

Pressing ENTR accepts the

new setting and returns to the

previous menu.

Pressing EXIT ignores the new

setting and returns to the

previous menu.

Press SET> to select the
analog output channel to
be configured. Then press
EDIT

to continue

DIAG AIO

CONC_OUT_2 OVERRANGE: ON

SET>

EDIT

EXIT

Figure 5-25:

Analog Output – Offset Adjustment