Pneumatic sensor assembly, Sample pressure, Izs option – Teledyne 6200T - Sulfides Analyzer User Manual

Page 259: 6200t total sulfur analyzer troubleshooting

Advertising
background image

6200T Total Sulfur Analyzer

Troubleshooting

Teledyen Analytical Instruments

259

Table 12-9: Example of HVPS Power Supply Outputs

If HVPS reading = 700 VDC

PIN PAIR

NOMINAL READING

KEY

5

6

7

8

9

10

11

1

2

3

4

1  2

70 VDC

2  3

70 VDC

3  4

70 VDC

4  5

70 VDC

5  6

70 VDC

6  7

70 VDC

7  8

70 VDC

8. Turn off the instrument power, and re-connect the PMT tube, and then re-assemble

the sensor.

If any faults are found in the test, the HVPS must be replaced. There are no user
serviceable parts inside the HVPS.

12.6.15. Pneumatic Sensor Assembly

The pressure/flow sensor circuit board, located behind the sensor assembly, can be
checked with a voltmeter using the following procedure, which assumes that the wiring
is intact and that the motherboard and the power supplies are operating properly.

 Measure the voltage across TP1 and TP2, it should be 10.0  0.25 V. If not,

the board may be faulty.

 Measure the voltage across capacitor C2; it should be 5.0 ± 0.25 V. If not,

the board may be faulty.

12.6.16. Sample Pressure

Measure the voltage across test points TP1 and TP4. With the sample pump
disconnected or turned off, this voltage should be 4500

 250 mV. With the pump

running, it should be about 0.2 V less as the sample pressure drops by about 1 in-Hg-A
from ambient pressure. If this voltage is significantly different, the pressure transducer
S2 or the board may be faulty. A leak in the sample system to vacuum may also cause
this voltage to be between about 0.6 and 4.5. Ensure that the front panel reading of the
sample pressure is at about 1 in-Hg-A less than ambient pressure.

12.6.17. IZS Option

The zero/span valves and IZS options need to be enabled in the software (contact the
factory on how to do this). Refer to Figure 3-19 and Figure 3-20 for a flow diagram with
zero/span valve or IZS option.

 Check for the physical presence of the valves or the IZS option.
 Check that a working perm-tube is installed in the IZS oven assembly.
 Check front panel for correct software configuration. When the instrument

is in SAMPLE mode, the front panel display should show CALS and CALZ
buttons in the second line of the display. The presence of the buttons
indicates that the option has been enabled in software. In addition, the IZS

Advertising