Teledyne 6200T - Sulfides Analyzer User Manual

Page 202

Advertising
background image

Calibration Procedures

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

202

Note

The CALIBRATE attribute must always be set to OFF for analyzers used in
US EPA controlled applications that have IZS option installed.

Calibration of instruments used in US EPA related applications should
only be performed using external sources of zero air and span gas with an
accuracy traceable to EPA or NIST standards and supplied through the
analyzer’s sample port (refer to Section 9.2).

The following example sets Sequence 2 to carry out a zero-span calibration every other
day starting at 01:00 on September 4, 2002, lasting 15 minutes. This sequence will start
0.5 hours later each day.

Table 9-4: Example Auto-Cal Sequence

MODE / ATTRIBUTE

VALUE

COMMENT

SEQUENCE

2

Define Sequence #2

MODE

ZERO-SPAN

Select Zero and Span Mode

TIMER ENABLE

ON

Enable the timer

STARTING DATE

Sept. 4, 2002

Start after Sept 4, 2002

STARTING TIME

01:00

First Span starts at 01:00

DELTA DAYS

2

Do Sequence #2 every other day

DELTA TIME

00:30

Do Sequence #2 0.5 h later each day

DURATION

15.0

Operate Span valve for 15 min

CALIBRATE

ON

The instrument will re-set the slope and offset
values for the SO

2

channel at the end of the AutoCal

sequence