Overview of operating modes – Teledyne 6200T - Sulfides Analyzer User Manual

Page 89

Advertising
background image

6200T Total Sulfur Analyzer

Overview of Operating Modes

Teledyne Analytical Instruments

89

4. OVERVIEW OF OPERATING MODES

To assist in navigating the analyzer’s software, a series of menu trees can be found in
Appendix A of this manual.

Note

Some control buttons on the touch screen do not appear if they are not
applicable to the menu that you’re in, the task that you are performing, the
command you are attempting to send, or to incorrect settings input by the
user. For example, the ENTR button may disappear if you input a setting
that is invalid or out of the allowable range for that parameter, such as
trying to set the 24-hour clock to 25:00:00. Once you adjust the setting to
an allowable value, the ENTR button will re-appear.

The 6200T software has a variety of operating modes. Most commonly, the analyzer
will be operating in SAMPLE mode. In this mode, a continuous read-out of the SO

2

concentration can be viewed on the front panel and output as an analog voltage from
rear panel terminals, calibrations can be performed, and TEST functions and
WARNING messages can be examined.
The second most important operating mode is SETUP mode. This mode is used for
performing certain configuration operations, such as for the DAS system, the reporting
ranges, or the serial (RS-232 / RS-485 / Ethernet) communication channels. The SET
UP mode is also used for performing various diagnostic tests during troubleshooting.

Figure 4-1: Front Panel Display

The Mode field of the front panel display indicates to the user which operating mode the
unit is currently running.
In addition to SAMPLE and SETUP, other modes available are presented in Table 4-1.

Table 4-1: Analyzer Operating Modes

MODE EXPLANATION

DIAG

One of the analyzer’s diagnostic modes is active (refer to Section 5.9).

Advertising