Milligrams per cubic meter (mg/m, Concentrations displayed in mg/m, And µg/m – Teledyne 6200T - Sulfides Analyzer User Manual
Page 104: Ppb x 1.34 = µg/m, Ppm x 1.34 = mg/m
Setup Menu
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
104
milligrams per cubic meter (mg/m
3
, MGM). Changing units affects all of the display,
analog outputs, COM port and DAS values for all reporting ranges regardless of the
analyzer’s range mode.
To change the concentration units:
SETUP X.X
CONC UNITS: PPM
PPM PPB
UGM
MGM
ENTER EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
CONC UNITS: PPB
PPM
PPB
UGM
MGM
ENTER
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
Select the preferred
concentration unit.
EXIT returns
to the main menu.
ENTR
accepts
the new unit,
EXIT
returns
to the SETUP
menu.
Figure 5-7: SETUP RNGE – Concentration Units Selection
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Concentrations displayed in mg/m
3
and µg/m
3
use 0°C and 760 Torr as
standard temperature and pressure (STP). Consult your local regulations
for the STP used by your agency. Here are the conversion factors from
volumetric to mass units used in the 6200T: SO
2
: ppb x 1.34 = µg/m
3
; ppm
x 1.34 = mg/m
3