6200T Total Sulfur Analyzer

Overview of Operating Modes

Teledyne Analytical Instruments

91

Table 4-2: Test Functions Defined

DISPLAY PARAMETER UNITS

DESCRIPTION

RANGE

RANGE

- -

RANGE1
RANGE2

PPB, PPM,

UGM & MGM

The Full Scale limit at which the reporting range of the analyzer’s
ANALOG OUTPUTS is currently set.
THIS IS NOT the Physical Range of the instrument. Refer to Section
5.4 for more information.
If DUAL or AUTO Range modes have been selected, two RANGE
functions will appear, one for each range.

STABIL

STABILITY mV

Standard deviation of SO

2

Concentration readings. Data points are

recorded every ten seconds. The calculation uses the last 25 data
points.

PRES

SAMPLE

PRESSURE

in-Hg-A

The current pressure of the sample gas as it enters the sample
chamber, measured between the SO

2

and Auto-Zero valves.

SAMP FL

SAMPLE FLOW

cm³/min

(cc/m)

The flow rate of the sample gas through the sample chamber. This
value is not measured but calculated from the sample pressure.

PMT

PMT Signal

mV

The raw output voltage of the PMT.

NORM PMT

NORMALIZED

PMT Signal

mV

The output voltage of the PMT after normalization for offset and
temperature/pressure compensation (if activated).

UV LAMP

Source UV Lamp

Intensity

mV

The output voltage of the UV reference detector.

LAMP

RATIO

UV Source lamp

ratio

%

The current output of the UV reference detector divided by the reading
stored in the CPU’s memory from the last time a UV Lamp calibration
was performed.

STR. LGT

Stray Light

ppb

The offset due to stray light recorded by the CPU during the last zero-
point calibration performed.

DRK PMT

Dark PMT

mV

The PMT output reading recorded the last time the UV source lamp
shutter was closed.

DRK LMP

Dark UV Source

Lamp

mV

The UV reference detector output reading recorded the last time the
UV source lamp shutter was closed.

SLOPE

SO

2

measurement

Slope

-

The sensitivity of the instrument as calculated during the last
calibration activity.

The slope parameter is used to set the span

calibration point of the analyzer.

OFFSET

SO

2

measurement

Offset

mV

The overall offset of the instrument as calculated during the last
calibration activity. The offset parameter is used to set the zero point
of the analyzer response.

HVPS

HVPS

V

The PMT high voltage power supply.

RCELL

TEMP

SAMPLE

CHAMBER TEMP

°C

The current temperature of the sample chamber.

BOX TEMP

BOX

TEMPERATURE

°C

The ambient temperature of the inside of the analyzer case.

PMT TEMP

PMT

TEMPERATURE

°C

The current temperature of the PMT.

IZS TEMP

1

IZS

TEMPERATURE

1

°C

The current temperature of the internal zero/span option. Only
appears when IZS option is enabled.

TEST

2

TEST SIGNAL

2

mV

Signal of a user-defined test function on output channel A4.

TIME

CLOCK TIME

hh:mm:ss

The current day time for DAS records and calibration events.

1

Only appears if Internal Gas Span Generator option is installed.

2

Only appears if analog output A3 is actively reporting a test function.