Remote Operation
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
184
To initialize the modem, access the COMMUNICATIONS Menu (refer to Figure 5-13),
then press:
SETUP X.X
COM1 MODEM INIT:AT Y
&D &H
<SET
SET>
EDIT
EXIT
SETUP X.X
COM1 INITIALIZE MODEM
<SET
SET> INIT
EXIT
SETUP X.X
COM1 MODE:0
SET>
EDIT
EXIT
SETUP X.X COM1 BAUD RATE:19200
<SET
SET>
EDIT
EXIT
EXIT
returns to the
Communications Menu.
SETUP X.X
INITIALIZING MODEM
<SET
SET> INIT
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID INET
COM1
COM2
EXIT
Select which
COM Port is
tested
Figure 8-5: COMM – Initialize the Modem
8.4. COM Port Password Security
In order to provide security for remote access of the 6200T, a LOGON feature can be
enabled to require a password before the instrument will accept commands. This is done
by turning on the SECURITY MODE (refer to Section 5.5). Once the SECURITY
MODE is enabled, the following items apply.
A password is required before the port will respond or pass on commands.
If the port is inactive for one hour, it will automatically logoff, which can
also be achieved with the LOGOFF command.
Three unsuccessful attempts to log on with an incorrect password will cause
subsequent logins to be disabled for 1 hour, even if the correct password is
used.
If not logged on, the only active command is the '?' request for the help
screen.
The following messages will be returned at logon:
LOGON SUCCESSFUL - Correct password given