5.9. SETUP – DIAG: Diagnostics Functions

The SETUP>MORE>DIAG menu provides a series of diagnostic functions whose
parameters are dependent on firmware revision (refer to Menu Tree, A-5, in Appendix
A). Table 5-3 describes the functions and provides a cross-reference to the details for
each in the remainder of this section. These functions can be used as tools in a variety of
troubleshooting and diagnostic procedures.

Table 5-3: 6200T Diagnostic (DIAG) Functions

DIAGNOSTIC FUNCTION AND MEANING

FRONT
PANEL

MODE

INDICATOR

SECTION

SIGNAL I/O:

Allows observation of all digital and analog

signals in the instrument. Allows certain digital signals such as
valves and heaters to be toggled ON and OFF.

DIAG I/O

5.9.1

ANALOG OUTPUT:

When entered, the analyzer performs an

analog output step test. This can be used to calibrate a chart
recorder or to test the analog output accuracy.

DIAG AOUT

5.9.2

ANALOG I/O CONFIGURATION:

Analog input/output

parameters are available for viewing and configuration.

DIAG AIO

5.9.3

OPTIC TEST:

When activated, the analyzer performs an optic

test, which turns on an LED located inside the sensor module
near the PMT (Fig. 10-15). This diagnostic tests the response
of the PMT without having to supply span gas.

DIAG OPTIC

5.9.4

ELECTRICAL TEST:

When activated, the analyzer performs

an electric test, which generates a current intended to simulate
the PMT output to verify the signal handling and conditioning of
the PMT preamp board.

DIAG ELEC

5.9.5

LAMP CALIBRATION:

The analyzer records the current

voltage output of the UV source reference detector. This value
is used by the CPU to calculate the lamp ration used in
determining the SO

2

concentration

DIAG LAMP

5.9.6

PRESSURE CALIBRATION:

The analyzer records the current

output of the sample gas pressure sensor. This value is used
by the CPU to compensate the SO

2

concentration when the

TPC feature is enabled.

DIAG PCAL

5.9.7

FLOW CALIBRATION:

This function is used to calibrate the

gas flow output signals of sample gas and ozone supply.
These settings are retained when exiting DIAG.

DIAG FCAL

5.9.8

TEST CHAN OUTPUT:

Configures the A4 analog output

channel.

DIAG TCHN

5.9.9