Troubleshooting

Model 6200T Total Sulfur Analyzer

270

Teledyne Analytical Instruments

not adequate).

10.

1. Set the analyzer display to show the signal I/O function, UVLAMP_SIGNAL (refer to

Section 12.1.3). UVLAMP_SIGNAL is function 33.

2. Slightly loosen the large brass thumbscrew located on the shutter housing (refer to

Figure 12-14) so that the lamp can be moved.

Figure 12-14. UV Lamp Adjustment

ATTENTION

COULD DAMAGE INSTRUMENT AND VOID WARRANTY

DO NOT grasp the UV lamp by its cap when changing its position -
always grasp the main body of the lamp (refer to Figure 12-13).
Inattention to this detail could twist and potentially disconnect the lamp’s
power supply wires.

3. While watching the UVLAMP_SIGNAL reading, slowly rotate the lamp or move it

back and forth vertically until the UVLAMP_SIGNAL reading is at its maximum.

4. Compare the UVLAMP_SIGNAL reading to the information in Table 12-10 and

follow the instructions there.

Table 12-10:

UV Lamp Signal Troubleshooting

UVLAMP_SIGNAL

ACTION TO BE TAKEN

3500mV±200mV. No

Action

Required

>

4900mV at any time.

Adjust the UV reference detector potentiometer (
Figure 12-15) until UVLAMP_SIGNAL reads approximately 3600mV before
continuing to adjust the lamp position.

>3700mV or < 3300mV

Adjust the UV reference detector potentiometer (
Figure 12-15) until UVLAMP_SIGNAL reads as close to 3500mV as possible.