Teledyne 6200T - Sulfides Analyzer User Manual
Page 270
Troubleshooting
Model 6200T Total Sulfur Analyzer
270
Teledyne Analytical Instruments
not adequate).
10.
1. Set the analyzer display to show the signal I/O function, UVLAMP_SIGNAL (refer to
Section 12.1.3). UVLAMP_SIGNAL is function 33.
2. Slightly loosen the large brass thumbscrew located on the shutter housing (refer to
Figure 12-14) so that the lamp can be moved.
Figure 12-14. UV Lamp Adjustment
ATTENTION
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
DO NOT grasp the UV lamp by its cap when changing its position -
always grasp the main body of the lamp (refer to Figure 12-13).
Inattention to this detail could twist and potentially disconnect the lamp’s
power supply wires.
3. While watching the UVLAMP_SIGNAL reading, slowly rotate the lamp or move it
back and forth vertically until the UVLAMP_SIGNAL reading is at its maximum.
4. Compare the UVLAMP_SIGNAL reading to the information in Table 12-10 and
follow the instructions there.
Table 12-10:
UV Lamp Signal Troubleshooting
UVLAMP_SIGNAL
ACTION TO BE TAKEN
3500mV±200mV. No
Action
Required
>
4900mV at any time.
Adjust the UV reference detector potentiometer (
Figure 12-15) until UVLAMP_SIGNAL reads approximately 3600mV before
continuing to adjust the lamp position.
>3700mV or < 3300mV
Adjust the UV reference detector potentiometer (
Figure 12-15) until UVLAMP_SIGNAL reads as close to 3500mV as possible.