6200t total sulfur analyzer setup menu, Single range mode (sngl), Teledyne analytical instruments 101 – Teledyne 6200T - Sulfides Analyzer User Manual
Page 101: Figure 5-4: setup rnge – single range mode
6200T Total Sulfur Analyzer
Setup Menu
Teledyne Analytical Instruments
101
5.4.3.1. Single Range mode (SNGL)
The default range mode for the analyzer is single range, in which all analog
concentration outputs are set to the same reporting range. This reporting range can be set
to any value between 0.1 ppb and 20,000 ppb.
While the two outputs always have the same reporting range, the span and scaling of
their electronic signals may also be configured for different differently (e.g., A1 = 0-10
V; A2 = 0-0.1 V).
To select SNGLE range mode and to set the upper limit of the range, press:
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL
IND AUTO
ENTR EXIT
EXIT
x 2 returns
to the main
SAMPLE
display
SETUP X.X
RANGE MODE:
SNGL
SNGL
IND AUTO
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
RANGE:
500.0 Conc
0 0
5
0 0 .0
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
Figure 5-4: SETUP RNGE – Single Range Mode