3.3.1.4. Current Loop Analog Outputs (Option 41) Setup

If your analyzer had this option installed at the factory, there are no further connections
to be made. The current loop option can be configured for any output range between 0
and 20 mA. Section 5.9.3.5 provides information on calibrating or adjusting these
outputs.
This section provides instructions for setting up the analog outputs for voltage and/or
current output. Figure 3-8 provides installation instructions and illustrates a sample
combination of one current output and two voltage outputs configuration.

 For current output install the Current Loop option PCA on J19, on J21 or

on J23 of the motherboard.

 For voltage output, install jumpers on J19, J21 and/or J23.

Following Figure 3-8 are instructions for converting current loop analog outputs to
standard 0-to-5 VDC outputs.

CAUTION – AVOID INVALIDATING WARRANTY

Servicing or handling of circuit components requires electrostatic
discharge protection, i.e. ESD grounding straps, mats and containers.
Failure to use ESD protection when working with electronic assemblies will
void the instrument warranty. Refer to Section 13 for more information on
preventing ESD damage.