13.4. Pneumatic Operation

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

It is important that the sample airflow system is leak-tight and not
pressurized over ambient pressure. Regular leak checks should be
performed on the analyzer as described in the maintenance schedule,
Table 11-1. Procedures for correctly performing leak checks can be found
in Section 11.3.6.

I

MPORTANT

Relative Pressure versus Absolute Pressure

In this manual vacuum readings are given in inches of mercury absolute
pressure (in-Hg-A), i.e. indicate an absolute pressure referenced against
zero (a perfect vacuum).

13.4.1. Sample Gas Flow

The Flow of gas through the 6200T UV Fluorescence SO

2

Analyzer is created by a

small internal pump that pulls air though the instrument.

V

ACUUM

M

ANI

F

O

L

D

FLOW
CONTROL
ASSY

EXHAUST TO OUTER

LAYER OF KICKER

SAMPLE
gas inlet

SAMPLE FILTER

Chassis

EXHAUST
gas outlet

KICKER EXHAUST

TO PUMP

HYDROCARBON

SCRUBBER

(KICKER)

UV
LAMP

PMT

SAMPLE

CHAMBER

FLOW

SENSOR

FLOW / PRESSURE

SENSOR PCA

SAMPLE

PRESSURE

SENSOR

PUMP

CRITICAL

FLOW

ORIFICE

Figure 13-10: Gas Flow and Location of Critical Flow Orifice