3. GETTING STARTED

This section addresses the procedures for unpacking the instrument and inspecting for
damage, presents clearance specifications for proper ventilation, introduces the
instrument layout, then presents the procedures for getting started: making electrical and
pneumatic connections, and conducting an initial calibration check.

3.1. Unpacking the 6200T Analyzer

CAUTION

GENERAL SAFETY HAZARD

To avoid personal injury, always use two persons to lift and carry the
6200T.

ATTENTION

COULD DAMAGE INSTRUMENT AND VOID WARRANTY

Printed Circuit Assemblies (PCAs) are sensitive to electro-static
discharges too small to be felt by the human nervous system. Failure to
use ESD protection when working with electronic assemblies will void
the instrument warranty. Refer to Section 13 for more information on
preventing ESD damage.

CAUTION

Do not operate this instrument until you’ve removed dust plugs from
SAMPLE and EXHAUST ports on the rear panel!

Note

Teledyne Analytical Instruments recommends that you store shipping
containers/materials for future use if/when the instrument should be
returned to the factory for repair and/or calibration service. See Warranty
section in this manual on page ii. Contact TAI at:

http://www.teledyne-

ai.com

f

or obtaining a Return Authorization.

Verify that there is no apparent external shipping damage. If damage has occurred,
please advise the shipper first, then Teledyne Analytical Instruments.
Included with your analyzer is a printed record of the final performance characterization
performed on your instrument at the factory. It is titled Final Test and Validation Data
Sheet
(P/N 04551). This record is an important quality assurance and calibration record
for this instrument. It should be placed in the quality records file for this instrument.
With no power to the unit, carefully remove the top cover of the analyzer and check for
internal shipping damage by carrying out the following steps: