Teledyne 6200T - Sulfides Analyzer User Manual
Page 36
Specifications
Model
6200T
Total Sulfur Analyzer
Teledyne Analytical Instruments
36
Parameter Description
Certifications
EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A,
ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A.
IEC 61010-1:90 + A1:92 + A2:95,
Approvals
North American:
cNEMKO (Canada): can/csa-c22.2 No. 61010-1-04
NEMKO-CCL (US): UL No. 61010-1 (2
nd
Edition)
US EPA: EQSA-0495-100
MCERTS: Sira MC 050067/04
EN14212, Approved, CE and Others
1
As defined by the USEPA.
2
Defined as twice the zero noise level by the USEPA.
Table 2-2: O2 Sensor Option Specifications
Parameter Description
Ranges
0-1% to 0-100% user selectable. Dual ranges and auto-ranging supported.
Zero Noise
1
<0.02% O
2
Lower Detectable Limit
2
<0.04%
O
2
Zero Drift (24 hours)
3
<± 0.02% O
2
Zero Drift (7 days)
<±- 0.05% O
2
Span Noise
1
<
±
0.05% O
2
Span Drift (7 days)
<± 0.1% O
2
Accuracy
(intrinsic error) <± 0.1% O
2
Linearity
<± 0.1 % O
2
Temp Coefficient
<± 0.05% O
2
/°C,
Rise and Fall Time
<60 seconds to 95%
1
As defined by the USEPA
2
Defined as twice the zero noise level by the USEPA
3
Note: zero drift is typically <± 0.1% O
2
during the first 24 hrs of operation
Table 2-3: CO
2
Sensor Option Specifications
Parameter Description
Ranges
0-1% to 0-20% user selectable. Dual ranges and auto-ranging supported.
Zero Noise
1
<0.02% CO
2
Lower Detectable Limit
2
<0.04%
CO
2
Zero Drift (24 hours)
<± 0.02% CO
2
Zero Drift (7 days)
<± 0.05% CO
2
Span Noise
1
<± 0.1% CO
2
Span Drift (7 days)
<± 0.1% CO
2
Accuracy
<± (0.02% CO
2
+ 2% of reading)