6200T Total Sulfur Analyzer

Getting Started

Teledyne Analytical Instruments

57

SO

2

Alarm 2 = 100 PPM

SO

2

Alarm 1 = 20 PPM

SO

2

Alarm 2 = 100 PPM

In this example, SO

2

Alarm 1 and SO

2

Alarm 1 will both be associated with the “Alarm

2” relay on the rear panel. This allows you to have multiple alarm levels for individual
concentrations.
A more likely configuration for this would be to put one concentration on the “Alarm 1”
relay and the other concentration on the “Alarm 2” relay.

SO

2

Alarm 1 = 20 PPM

SO

2

Alarm 2 = Disabled

SO

2

Alarm 1 = Disabled

SO

2

Alarm 2 = 100 PPM

“Alarm 4” Relay

This relay is connected to the “range bit”. If the instrument is configured for “Auto
Range” and the reading goes up into the high range, it will turn this relay on.

3.3.1.8. Connecting the Communications Interfaces

The T-Series analyzers are equipped with connectors for remote communications
interfaces: Ethernet, USB, RS-232, RS-232 Multidrop and RS-485. In addition to
using the appropriate cables, each type of communication method, must be configured
using the SETUP>COMM menu, Section 6. Although Ethernet is DHCP-enabled by
default, it can also be configured manually (Section 6.5.1) to set up a static IP address,
which is the recommended setting when operating the instrument via Ethernet.

Ethernet Connection

For network or Internet communication with the analyzer, connect an Ethernet cable
from the analyzer’s rear panel Ethernet interface connector to an Ethernet port. Please
refer to Section 6.5 for a description of the default configuration and setup instructions.

Configuration: Section 6.5

 manual configuration: Section 6.5.1
 automatic configuration (default): Section 6.5.2

USB Connection

For direct communication between the analyzer and a PC, connect a USB cable between
the analyzer and desktop or laptop USB ports, and ensure that their baud rates match
(Section 6.2.2).

Note

If this option is installed, the COM2 port cannot be used for anything
other than Multidrop communication.

Configuration: Section 6.2.2