9.4. Manual Calibration with Zero/Span Valves

Zero and Span calibrations using the Zero/Span Valve option are similar to that
described in Section 7.2, except that:

 Zero air and span gas are supplied to the analyzer through the zero gas and

span gas inlets rather than through the sample inlet.

 The zero and cal operations are initiated directly and independently with

dedicated buttons (CALZ and CALS)

STEP ONE: Connect the sources of zero air and span gas to the respective ports on the
rear panel (refer to Figure 3-1) as shown below.

External Zero

Air Scrubber

Filter

Source of

SAMPLE Gas

VENT if input is pressurized

MODEL 700

Gas Dilution Calibrator

(with O

3

generator option)

VENT

SAMPLE

ZERO AIR

EXHAUST

SPAN 1

Calibrated

SO

2

Gas

(At high

concentration)

Needle valve

to control flow

VE

NT

MODEL 701

Zero Air

Generator

Chassis

Figure 9-5: Setup for Manual Calibration with Z/S Valve Option Installed (Step 1)