Teledyne 6200T - Sulfides Analyzer User Manual

Page 234

Advertising
background image

Maintenance

Model 6200T Total Sulfur Analyzer

234

Teledyne Analytical Instruments

11.3.8.1. Checking the Scrubber for Leaks

Leaks in the outer tubing of the scrubber can be found using the procedure described in
Section 11.3.6. Use the following method to determine if a leak exists in the inner
tubing of the scrubber.
This procedure requires a pressurized source of air (chemical composition is
unimportant) capable of supplying up to 15 psiA and a leak checking fixture such as the
one illustrated in Figure 11-3.

Vacuum/Pressure

Gauge

Manual Shut-Off

Valve

Needle Valve

FROM PUMP or

PRESSURIZED

AIR SOURCE

TO SCRUBBER

Figure 11-3:

Simple Leak Check Fixture

1. Turn off the analyzer.

2. Disconnect the pneumatic tubing attached to both ends of the scrubber’s inner

tubing.

3. One end is connected to the sample particulate filter assembly and the other end

is connected to the reaction cell assembly.

4. Both ends are made of the 1/8" black Teflon tubing.

5. Cap one end of the hydrocarbon scrubber.

6. Attach the pressurized air source to the other end of the scrubber inner tubing

with the leak check fixture in line.

Pump

or

Pressurized Air

Source

Leak Check

Fixture

Scrubber

Cap

Figure 11-4:

Hydrocarbon Scrubber Leak Check Setup

7. Use the needle valve to adjust the air input until the gauge reads 15 psiA.

ATTENTION

COULD DAMAGE INSTRUMENT AND VOID WARRANTY

Do not exceed a pressure of more than 15 psia.

Do not pull the vacuum through the scrubber.

8. Close the shut-off valve.

Advertising