Specifications, approvals & compliance, Specifications and approvals, 6200t total sulfur analyzer specifications – Teledyne 6200T - Sulfides Analyzer User Manual
Page 35
6200T Total Sulfur Analyzer
Specifications
Teledyne Analytical Instruments
35
2. SPECIFICATIONS, APPROVALS & COMPLIANCE
This section presents specifications for the 6200T analyzer and the O
2
and CO
2
sensor
options, Agency approvals, EPA equivalency designation, and CE mark compliance.
2.1. Specifications and Approvals
Table 2-1 6200T Basic Unit Specifications
Parameter Description
Ranges
(Physical Analog Output)
Min: 0-50 ppb Full Scale
Max: 0-20,000 ppb Full Scale (selectable, dual ranges and auto ranging supported)
Measurement Units
ppb, ppm, µg/m
3
, mg/m
3
(selectable)
Zero Noise
1
<
0.2 ppb (RMS)
Span Noise
1
<
0.5% of reading, above 50 ppb
Lower Detectable Limit
2
0.4
ppb
Zero Drift)
<
0.5 ppb/24 hours
Span Drift
<
0.5% of full scale/24 hours
Lag Time
1
20
seconds
Rise/Fall Time
1
<100 sec to 95%
Linearity
1% of full scale
Precision
1
0.5% of reading above 50 ppb
Sample Flow Rate
650 cm
3
/min. ±10%
Power Requirements
100V-120V, 220V-240 V, 50/60 Hz
Analog Output Ranges
10 V, 5 V, 1 V, 0.1 V (selectable)
Recorder Offset
± 10 %
Standard I/O
1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital outputs
6 opto-isolated digital inputs
4 analog outputs
Optional I/O
1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs
Multidrop RS232
3 4-20mA current outputs
Environmental Installation
category (over-voltage category) II; Pollution degree 2
Operating Temperature
Range
5 - 40
o
C (with EPA Equivalency)
Humidity Range
0 - 95% RH, non-condensing
Dimensions HxWxD
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
Weight
31 lbs (14 kg)
35.7 lbs (16 kg) with internal pump