Specifications, approvals & compliance, Specifications and approvals, 6200t total sulfur analyzer specifications – Teledyne 6200T - Sulfides Analyzer User Manual

Page 35

Advertising
background image

6200T Total Sulfur Analyzer

Specifications

Teledyne Analytical Instruments

35

2. SPECIFICATIONS, APPROVALS & COMPLIANCE

This section presents specifications for the 6200T analyzer and the O

2

and CO

2

sensor

options, Agency approvals, EPA equivalency designation, and CE mark compliance.

2.1. Specifications and Approvals

Table 2-1 6200T Basic Unit Specifications

Parameter Description

Ranges
(Physical Analog Output)

Min: 0-50 ppb Full Scale
Max: 0-20,000 ppb Full Scale (selectable, dual ranges and auto ranging supported)

Measurement Units

ppb, ppm, µg/m

3

, mg/m

3

(selectable)

Zero Noise

1

<

0.2 ppb (RMS)

Span Noise

1

<

0.5% of reading, above 50 ppb

Lower Detectable Limit

2

0.4

ppb

Zero Drift)

<

0.5 ppb/24 hours

Span Drift

<

0.5% of full scale/24 hours

Lag Time

1

20

seconds

Rise/Fall Time

1

<100 sec to 95%

Linearity

1% of full scale

Precision

1

0.5% of reading above 50 ppb

Sample Flow Rate

650 cm

3

/min. ±10%

Power Requirements

100V-120V, 220V-240 V, 50/60 Hz

Analog Output Ranges

10 V, 5 V, 1 V, 0.1 V (selectable)

Recorder Offset

± 10 %

Standard I/O

1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital outputs
6 opto-isolated digital inputs
4 analog outputs

Optional I/O

1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs
Multidrop RS232
3 4-20mA current outputs

Environmental Installation

category (over-voltage category) II; Pollution degree 2

Operating Temperature
Range

5 - 40

o

C (with EPA Equivalency)

Humidity Range

0 - 95% RH, non-condensing

Dimensions HxWxD

7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)

Weight

31 lbs (14 kg)
35.7 lbs (16 kg) with internal pump

Advertising