Teledyne 6200T - Sulfides Analyzer User Manual

Page 190

Advertising
background image

Calibration Procedures

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

190

OR

SAMPLE

EXHAUST

MODEL 701

Zero Air

Generator

VE

N

T

Calibrated

SO

2

GAS

(At high

concentration)

MODEL 700

Gas Dilution

Calibrator

(with Ozone

Bench Option)

Source of

SAMPLE Gas

(Remove

during

calibration)

SAMPLE

EXHAUST

VE

N

T

Calibrated

SO

2

GAS

(At high

concentration)

MODEL 701

Zero Air

Generator

Source of

SAMPLE Gas

(Remove

during

calibration)

3-way

Valve

Needle valve

to control

flow

Chassis

Chassis

Figure 9-1: Setup for Manual Calibration without Z/S valve or IZS Option (Step 1)

STEP TWO: Set the expected

SO

2

span gas concentrations. In this example the

instrument is set for single (SNGL) range mode with a reporting range span of 500 ppb.

Advertising