Signal i/o, 6200t total sulfur analyzer setup menu – Teledyne 6200T - Sulfides Analyzer User Manual

Page 117

Advertising
background image

6200T Total Sulfur Analyzer

Setup Menu

Teledyne Analytical Instruments

117

5.9.1. Signal I/O

The signal I/O diagnostic mode allows a user to review and change the digital and
analog input/output functions of the analyzer. Refer to Appendix A-4 for a complete list
of the parameters available for review under this menu.

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

Any changes of signal I/O settings will remain in effect only until the
signal I/O menu is exited. Exceptions are the ozone generator override
and the flow sensor calibration, which remain as entered when exiting.

Access the signal I/O test mode from the DIAG Menu (refer to Figure 5-16), then press:

EXAMPLE

DIAG

SIGNAL I / O

PREV NEXT

JUMP

ENTR

EXIT

DIAG I / O

0) EXT_ZERO_CAL=OFF

PREV NEXT

JUMP

PRNT EXIT

Press JUMP to go

directly to a specific

signal

See Appendix A-4 for

a complete list of

available SIGNALS

DIAG I / O

JUMP TO: 12

1 2

ENTR

EXIT

EXAMPLE:

Enter 12 to Jump to

12)

ST_SYSTEM_OK=ON

DIAG I / O

12)

ST_SYSTEM_OK = ON

PREV NEXT

JUMP

ON

PRNT

EXIT

Exit to return

to the

DIAG

menu

Press NEXT & PREV to

move between signal

types.

Pressing the PRNT button will send a formatted

printout to the serial port and can be captured

with a computer or other output device.

Toggle ON/(OFF) button to

change status.

Figure 5-17:

DIAG – Signal I/O Menu