5.9.1. Signal I/O
The signal I/O diagnostic mode allows a user to review and change the digital and
analog input/output functions of the analyzer. Refer to Appendix A-4 for a complete list
of the parameters available for review under this menu.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Any changes of signal I/O settings will remain in effect only until the
signal I/O menu is exited. Exceptions are the ozone generator override
and the flow sensor calibration, which remain as entered when exiting.
Access the signal I/O test mode from the DIAG Menu (refer to Figure 5-16), then press:
EXAMPLE
DIAG
SIGNAL I / O
PREV NEXT
JUMP
ENTR
EXIT
DIAG I / O
0) EXT_ZERO_CAL=OFF
PREV NEXT
JUMP
PRNT EXIT
Press JUMP to go
directly to a specific
signal
See Appendix A-4 for
a complete list of
available SIGNALS
DIAG I / O
JUMP TO: 12
1 2
ENTR
EXIT
EXAMPLE:
Enter 12 to Jump to
12)
ST_SYSTEM_OK=ON
DIAG I / O
12)
ST_SYSTEM_OK = ON
PREV NEXT
JUMP
ON
PRNT
EXIT
Exit to return
to the
DIAG
menu
Press NEXT & PREV to
move between signal
types.
Pressing the PRNT button will send a formatted
printout to the serial port and can be captured
with a computer or other output device.
Toggle ON/(OFF) button to
change status.
Figure 5-17:
DIAG – Signal I/O Menu