Flow calibration – Teledyne 6200T - Sulfides Analyzer User Manual

Page 136

Advertising
background image

Setup Menu

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

136

5.9.8. Flow Calibration

The flow calibration allows the user to adjust the values of the sample flow rates as they
are displayed on the front panel and reported through COM ports to match the actual
flow rate measured at the sample inlet. This does not change the hardware measurement
of the flow sensors, only the software calculated values.
To carry out this adjustment, connect an external, sufficiently accurate flow meter to the
SAMPLE inlet (refer to Section 11 for more details).
Once the flow meter is attached and is measuring actual gas flow, access the Signal I/O
from the DIAG Menu (refer to Figure 5-16), then press:

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

Repeat Pressing NEXT until . .

.

DIAG

FLOW CALIBRATION

PREV NEXT

ENTR

EXIT

DIAG FCAL

ACTUAL FLOW: 607 CC / M


0 6 0 7 ENTR EXIT

ENTR accepts the

new value

EXIT

ignores the new

value

Adjust these values until the

displayed flow rate equals the

flow rate being measured by the

independent flow meter.

Figure 5-34:

DIAG – Flow Calibration