References, 220 teledyne analytical instruments, Figure 10-1: dynamic multipoint span calibration – Teledyne 6200T - Sulfides Analyzer User Manual
Page 220
EPA Protocol Calibration
Model 6200T Total Sulfur Analyzer
220
Teledyne Analytical Instruments
Wait until
STABIL
falls
below 0.5 ppb.
This may take
several minutes.
SAMPLE
STABIL=X.XXX PPB
SO2 =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
Record the SO
2
reading as
displayed on the
instrument’s front
panel
ACTION:
Allow calibration gas diluted to proper concentration for
Midpoint N to enter the sample port
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST
TST
>
CAL
SETUP
SAMPLE
STABIL=X.XXX PPB
SO2 =XXX.X
< TST TST >
CAL
SETUP
Set the Display to show the
STABIL
test function.
This function calculates the
stability of the SO
2
measurement
SPAN CAL M
RANGE = 500.0 PPB
SO
2
XXX.X
< TST TST >
ZERO
SPAN
CONC
EXIT
ACTION:
Allow Calibration Gas diluted to
proper concentration for
Midpoint N+1 to enter the sample
port
Press EXIT to
Return to the
Main SAMPLE
Display
Figure 10-1:
Dynamic Multipoint Span Calibration
10.6. Special Calibration Requirements for Dual Range or Auto
Range
If Dual Range or Auto Range is selected, then calibration for Range1 and Range2
should be performed separately.
For zero and span point calibration, follow the procedure described in Section 9.2.
Repeat the procedure for both the HIGH and LOW Ranges.
10.7. References
1. Environmental Protection Agency, Title 40, Code of Federal Regulations, Part 50,
Appendix A, Section 10.3.