Getting Started

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

54

Table 3-7: Status Output Signals

REAR PANEL

LABEL

STATUS

DEFINITION

CONDITION

1

SYSTEM OK

ON if no faults are present.

2

CONC VALID

OFF any time the HOLD OFF feature is active, such as during calibration or when
other faults exist possibly invalidating the current concentration measurement
(example: sample flow rate is outside of acceptable limits).
ON if concentration measurement is valid.

3

HIGH RANGE

ON if unit is in high range of either the DUAL or Auto range modes.

4

ZERO CAL

ON whenever the instrument’s ZERO point is being calibrated.

5

SPAN CAL

ON whenever the instrument’s SPAN point is being calibrated.

6

DIAG MODE

ON whenever the instrument is in DIAGNOSTIC mode

7&8

SPARE

D

EMITTER BUS

The emitters of the transistors on pins 1-8 are bussed together.

SPARE

+

DC POWER

+ 5 VDC, 300 mA source (combined rating with Control Output, if used).

Digital Ground

The ground level from the analyzer’s internal DC power supplies

3.3.1.6. Connecting the Control Inputs

If you wish to use the analyzer to remotely activate the zero and span calibration modes,
several digital control inputs are provided through a 10-pin connector labeled
CONTROL IN on the analyzer’s rear panel.
There are two methods for energizing the control inputs. The internal +5V available
from the pin labeled “+” is the most convenient method. However, if full isolation is
required, an external 5 VDC power supply should be used.