Setup – pass: password protection – Teledyne 6200T - Sulfides Analyzer User Manual

Page 106

Advertising
background image

Setup Menu

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

106

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

Once the above settings have been entered, the instrument needs to be
recalibrated using one of the methods discussed in Section 9.

5.5. SETUP – PASS: Password Protection

The menu system provides password protection of the calibration and setup functions to
prevent unauthorized adjustments. When the passwords have been enabled in the PASS
menu item, the system will prompt the user for a password anytime a password-
protected function (e.g., SETUP) is selected. This allows normal operation of the
instrument, but requires the password (101) to access to the menus under SETUP. When
PASSWORD is disabled (SETUP>OFF), any operator can enter the Primary Setup
(SETUP) and Secondary Setup (SETUP>MORE) menus. Whether PASSWORD is
enabled or disabled, a password (default 818) is required to enter the VARS or DIAG
menus in the SETUP>MORE menu.

Table 5-1: Password Levels

PASSWORD

LEVEL

MENU ACCESS ALLOWED

Null (000)

Operation

All functions of the main menu (top level, or Primary, menu)

101

Configuration/Maintenance

Access to Primary and Secondary SETUP Menus when PASSWORD is
enabled

818

Configuration/Maintenance Access to Secondary SETUP Submenus VARS and DIAG whether

PASSWORD is enabled or disabled.