6200t total sulfur analyzer calibration procedures, Activate co, Sensor stability function – Teledyne 6200T - Sulfides Analyzer User Manual
Page 211
6200T Total Sulfur Analyzer
Calibration Procedures
Teledyne Analytical Instruments
211
9.10.2.3. Activate CO
2
Sensor Stability Function
To change the stability test function from SO
2
concentration to the CO
2
sensor output,
press:
Press
3
times to return
to
menu.
SETUP X.X
SO2 CO2 O2
ENTR
SAMPLE RANGE=50.0 PPM SO2= XX.XX
< TST TST > CAL
SETUP X.X
CFG DAS RNGE PASS CLK
EXIT
SETUP X.X
COMM
DIAG ALRM
EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV
JUMP
EDIT PRNT EXIT
SETUP X.X
<PREV NEXT> JUMP
PRNT EXIT
SETUP X.X
SO2
O2
ENTR EXIT
Continue pressing
until ...
SETUP X.X
EXIT
CO
2
and O
2
options only
appear if
associated
sensors are
installed.
Figure 9-19:
Activate CO
2
Sensor Stability Function
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Use the same procedure to reset the STB test function to SO
2
when the
CO
2
calibration procedure is complete.