EPA Protocol Calibration

Model 6200T Total Sulfur Analyzer

214

Teledyne Analytical Instruments

Table 10-1:

Activity Matrix for Calibration Equipment & Supplies

EQUIPMENT &

SUPPLIES

ACCEPTANCE LIMITS

FREQUENCY AND METHOD

OF MEASUREMENT

ACTION IF

REQUIREMENTS ARE NOT

MET

Recorder

Compatible with output
signal of analyzer; min.
chart width of 150 mm (6 in)
is recommended

Check upon receipt

Return equipment to supplier

Sample line and

manifold

Constructed of PTFE or
glass

Check upon receipt

Return equipment to supplier

Calibration equipment Meets guidelines of

reference 1 and Section
2.3.2 (Q.A. Handbook)

Refer to Section 2.3.9 (Q.A.
Handbook)

Return equipment/ supplies
to supplier or take corrective
action

Working standard SO

2

cylinder gas or SO

2

permeation tube

Traceable to NIST-SRM
meets limits in traceability
protocol for accuracy and
stability (refer to Section
2.0.7, Q.A. Handbook)

Analyzed against NIST-SRM;
refer to protocol in Section
2.0.7, Q.A. Handbook

Obtain new working
standard and check for
traceability

Zero air

Clean dry ambient air, free
of contaminants that cause
detectable response with
the SO

2

analyzer.

Refer to Section 2.9.2 (Q.A.
Handbook)

Obtain air from another
source or regenerate.

Record form

Develop standard forms

N/A

Revise forms as appropriate

Audit equipment

Must not be the same as
used for calibration

System must be checked out
against known standards

Locate problem and correct
or return to supplier

Table 10-2:

Activity Matrix for Calibration Procedure

EQUIPMENT &

SUPPLIES

ACCEPTANCE LIMITS

FREQUENCY AND METHOD

OF MEASUREMENT

ACTION IF

REQUIREMENTS ARE NOT

MET

Calibration gases

NIST traceable

Assayed against an NIST-
SRM semi-annually, Sec.
2.0.7, (Q.A. Handbook)

Working gas standard is
unstable, and/or
measurement method is out
of control; take corrective
action such as obtaining new
calibration gas.

Dilution gas

Zero air, free of
contaminants

Refer to Section 2.9.2 (Q.A.
Manual)

Return to supplier or take
appropriate action with
generation system

Multi-point calibration

Use calibration
procedure in Subsec. 2.2
(Q.A. Handbook); also
Federal Register

Perform at least once every
quarter or anytime a level
span check indicates a
discrepancy, or after
maintenance which may
affect the calibration; Subsec
2.5 (Q.A. Manual)

Repeat the calibration

10.1.2. Data Recording Device

Either a strip chart recorder, data acquisition system, digital data acquisition system
should be used to record the data from the Mode; 6200T RS-232 port or analog outputs.
If analog readings are being used, the response of that system should be checked against