Optic test – Teledyne 6200T - Sulfides Analyzer User Manual

Page 132

Advertising
background image

Setup Menu

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

132

5.9.4. Optic Test

The optic test function tests the response of the PMT sensor by turning on an LED
located in the cooling block of the PMT (refer to
Figure 13-18). The analyzer uses the light emitted from the LED to test its photo-
electronic subsystem, including the PMT and the current to voltage converter on the pre-
amplifier board. To ensure that the analyzer measures only the light coming from the
LED, the analyzer should be supplied with zero air. The optic test should produce a
PMT signal of about 2000±1000 mV.

Access the Optic Test from the DIAG Menu (refer to Figure 5-16), then press:

DIAG

SIGNAL I / O


PREV NEXT

JUMP

ENTR

EXIT

DIAG OPTIC

RANGE = 500.000 PPB SO2=XXX.X

<TST TST>

EXIT

Press NEXT until…

DIAG

OPTIC TEST

PREV NEXT

ENTR

EXIT

Press TST until…

DIAG ELEC

PMT = 2751 MV

SO2=XXX.X

<TST TST>

EXIT

While the optic test is

activated, PMT should be

2000 mV ± 1000 mV

Figure 5-30:

DIAG – Optic Test

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

This is a coarse test for functionality and not an accurate calibration tool.
The resulting PMT signal can vary significantly over time and also
changes with low-level calibration.

Advertising