Teledyne 6200T - Sulfides Analyzer User Manual
Page 61
6200T Total Sulfur Analyzer
Getting Started
Teledyne Analytical Instruments
61
3. Close the instrument.
4. Referring to Figure 3-15 use straight-through DB9 male DB9 female cables
to interconnect the host RS232 port to the first analyzer’s RS232 port; then
from the first analyzer’s COM2 port to the second analyzer’s RS232 port; from
the second analyzer’s COM2 port to the third analyzer’s RS232 port, etc.,
connecting in this fashion up to eight analyzers, subject to the distance
limitations of the RS-232 standard.
5. BEFORE communicating from the host, power on the instruments and check
that the Machine ID code is unique for each (Section 5.7.1). On the front panel
menu, use SETUP>MORE>COMM>ID. Note that the default ID is typically
the model number; to change the 4-digit identification number, press the button
of the digit to be changed).
Note
Teledyne Analytical Instruments recommends setting up the first link,
between the Host and the first analyzer, and testing it before setting up
the rest of the chain.
Analyzer
Analyzer
Analyzer
Last Analyzer
Female DB9
Male DB9
RS-232
COM2
RS-232
COM2
RS-232
COM2
RS-232
COM2
Host
RS-232 port
Ensure jumper is
installed between
JP2 pins 21
22
in last instrument
of multidrop chain.
Figure 3-15:
RS-232-Multidrop PCA Host
/
Analyzer Interconnect Diagram
RS-485 Connection
As delivered from the factory, COM2 is configured for RS-232 communications. This
port can be reconfigured for operation as a non-isolated, half-duplex RS-485 port. Using
COM2 for RS-485 communication disables the USB port. To reconfigure this port for
RS-485 communication, please contact the factory.