Test channel output, 6200t total sulfur analyzer setup menu – Teledyne 6200T - Sulfides Analyzer User Manual
Page 137
6200T Total Sulfur Analyzer
Setup Menu
Teledyne Analytical Instruments
137
5.9.9. Test Channel Output
When activated, output channel A3 can be used in the standard configuration to report
one of the test functions viewable from the SAMPLE mode display.
To activate the A3 channel and select a test function, access the Signal I/O from the
DIAG Menu (refer to Figure 5-16), then press:
DIAG TCHN
TEST CHANNEL: NONE
NEXT
ENTR
EXIT
DIAG
TEST CHAN OUTPUT
PREV
NEXT
ENTR
EXIT
Continue to press NEXT until …
DIAG TCHN
TEST CHANNEL: PMT READING
PREV
NEXT
ENTR
EXIT
Press PREV or NEXT
to move through the
list of available
parameters
(Table 6-9)
Press EXIT to
return to the
DIAG
menu
Press ENTR to
select the displayed
parameter and to
activate the test
channel.
DIAG
SIGNAL I / O
PREV
NEXT
ENTR
EXIT
Figure 5-35:
DIAG – Test Channel Output
Table 5-9:
Test Parameters Available for Analog Output A3 (standard
configuration)
Test Channel
Test parameter range
NONE
Test channel is turned off
PMT READING
0-5000 mV
UV READING
0-5000 mV
SAMPLE PRESSURE
0-40 in-Hg-A
SAMPLE FLOW
0-1000 cm³/min
RCELL TEMP
0-70° C
CHASSIS TEMP
0-70° C
IZS TEMP
0-70° C
PMT TEMP
0-50° C
HVPS VOLTAGE
0-5000 V