Teledyne 6200T - Sulfides Analyzer User Manual
Page 342
Instrument Test and Calibration Record
Model 6200T Total Sulfur Analyzer
342
Teledyne Analytical Instruments
Table 17-2.
For test and calibration information with CO2, refer to Table 17-3.
Table 17-1.
Final Test and Calibration Values for 6200T
TEST Parameters
Observed Value
Units
Acceptable Value
RANGE
PPB
50 - 20,000
STABIL
PPB
0.0 - 2
PRESS
“ HG
24 - 35
SAMP FL
CC / MIN 500 - 700 w/CO
2
PMT
mV
0 - 5000
UV LAMP
mV
3500 - 4000
STR. LGT
PPB
< 60
DRK PMT
MV
< 50
DRK LMP
MV
< 50
SLOPE
1.0
± 0
.3
OFFSET
MV
< 100
HVPS
V
400 - 900 constant
DCPS
MV
2500 +/- 200
RCELL TEMP
o
C
50 +/- 1
BOX TEMP
o
C 8-50
PMT TEMP
o
C
7.9 +/- 1
IZS TEMP
o
C 50
+/-.3
Electric Test
PMT Volts
MV
1000 +/-200
TS Conc
PPB
500 +/- 100
Optic Test
PMT Volts
MV
1000 +/- 200
TS Conc
PPB
500 +/- 100