Instrument Test and Calibration Record

Model 6200T Total Sulfur Analyzer

342

Teledyne Analytical Instruments

Table 17-2.
For test and calibration information with CO2, refer to Table 17-3.

Table 17-1.

Final Test and Calibration Values for 6200T

TEST Parameters

Observed Value

Units

Acceptable Value

RANGE

PPB

50 - 20,000

STABIL

PPB

0.0 - 2

PRESS

“ HG

24 - 35

SAMP FL

CC / MIN 500 - 700 w/CO

2

PMT

mV

0 - 5000

UV LAMP

mV

3500 - 4000

STR. LGT

PPB

< 60

DRK PMT

MV

< 50

DRK LMP

MV

< 50

SLOPE

1.0

± 0

.3

OFFSET

MV

< 100

HVPS

V

400 - 900 constant

DCPS

MV

2500 +/- 200

RCELL TEMP

o

C

50 +/- 1

BOX TEMP

o

C 8-50

PMT TEMP

o

C

7.9 +/- 1

IZS TEMP

o

C 50

+/-.3

Electric Test

PMT Volts

MV

1000 +/-200

TS Conc

PPB

500 +/- 100

Optic Test

PMT Volts

MV

1000 +/- 200

TS Conc

PPB

500 +/- 100