background image

6200T Total Sulfur Analyzer

Setup Menu

Teledyne Analytical Instruments

97

5. SETUP MENU

The SETUP menu is used to set instrument parameters for performing configuration,
calibration, reporting and diagnostics operations according to user needs.

5.1. SETUP – CFG: Configuration Information

Pressing the CFG button displays the instrument configuration information. This display
lists the analyzer model, serial number, firmware revision, software library revision,
CPU type and other information. Use this information to identify the software and
hardware when contacting customer service. Special instrument or software features or
installed options may also be listed here.

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818


8 1 8

ENTR

EXIT

SAMPLE

PRIMARY SETUP MENU

CFG

DAS RNGE PASS CLK MORE EXIT

SAMPLE

T100 SO2 ANALYZER

NEXT

PREV

EXIT

Press EXIT at

any time to

return to

SETUP menu

Press EXIT at

any time to

return to the

SAMPLE display

Press NEXT of PREV to move back

and forth through the following list

of Configuration information:

MODEL NAME
SERIAL NUMBER
SOFTWARE REVISION
LIBRARY REVISION

iCHIP SOFTWARE REVISION

1

HESSEN PROTOCOL REVISION

1

ACTIVE SPECIAL SOFTWARE

OPTIONS

1

CPU TYPE
DATE FACTORY CONFIGURATION

SAVED

1

Only appears if relevant option of Feature is active.

Figure 5-1: SETUP – Configuration Information

5.2. SETUP – ACAL: Automatic Calibration Option

The menu button for this option appears only when the instrument has the zero span
and/or IZS options. See Section 9.8 for details.

5.3. SETUP – DAS: Internal Data Acquisition System

Use the SETUP>DAS menu to capture and record data. Refer to Section 6.6 for
configuration and operation details.

5.4. SETUP – RNGE: Analog Output Reporting Range

Configuration

Use the SETUP>RNGE menu to configure output reporting ranges, including scaled
reporting ranges to handle data resolution challenges. This section describes
configuration for Single, Dual, and Auto Range modes.