Instrument layout, Ventilation clearance, Front panel – Teledyne 6200T - Sulfides Analyzer User Manual

Page 40

Advertising
background image

Getting Started

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

40

1. Remove the locking screw located in the top, center of the Front panel;

2. Remove the two flat head, Phillips screws on the sides of the instrument;

3. Slide the cover backwards until it clears the analyzer’s front bezel, and;

4. Lift the cover straight up.

5. Inspect the interior of the instrument to ensure that all circuit boards and other

components are in good shape and properly seated.

6. Check the connectors of the various internal wiring harnesses and pneumatic

hoses to ensure that they are firmly and properly seated.

7. Verify that all of the optional hardware ordered with the unit has been installed.

These are listed on the paperwork accompanying the analyzer.

WARNING

ELECTRICAL SHOCK HAZARD

Never disconnect PCAs, wiring harnesses or electronic subassemblies
while under power.

3.1.1. Ventilation Clearance

Whether the analyzer /converter are set up on a bench or installed into an instrument
rack, be sure to leave sufficient ventilation clearance.

Table 3-1: Ventilation Clearance

AREA

MINIMUM REQUIRED

CLEARANCE

Back of the instrument

4 in.

Sides of the instrument

1 in.

Above and below the

instrument

1 in.

Various rack mount kits are available for this analyzer. Refer to Section 1.4 of this
manual for more information.

3.2. Instrument Layout

Instrument layout includes front panel and display, rear panel connectors, and internal
chassis layout.

3.2.1. Front Panel

Figure 3-1 shows the analyzer’s front panel layout, followed by a close-up of the
display screen in
Figure 3-2, which is described in Table 3-2. The two USB ports on the front panel are
provided for the connection of peripheral devices:

 Plug-in mouse (not included) to be used as an alternative to the touchscreen

interface