DAS and AICOM

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

176

Figure 7-15:

DAS Configuration Through a Terminal Emulation Program

Both procedures are best started by downloading the default DAS configuration, getting
familiar with its command structure and syntax conventions, and then altering a copy of
the original file offline before uploading the new configuration.

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

Whereas the editing, adding and deleting of DAS channels and
parameters of one channel through the front-panel control buttons can
be done without affecting the other channels, uploading a DAS
configuration script to the analyzer through its communication ports will
erase all data, parameters and channels by replacing them with the new
DAS configuration. Backup of data and the original DAS configuration is
advised before attempting any DAS changes.