Connecting analog outputs – Teledyne 6200T - Sulfides Analyzer User Manual

Page 50

Advertising
background image

Getting Started

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

50

Table 3-5: Analog Input Pin Assignments

PIN DESCRIPTION

DAS

PARAMETER

1

1

Analog input # 1

AIN 1

2

Analog input # 2

AIN 2

3

Analog input # 3

AIN 3

4

Analog input # 4

AIN 4

5

Analog input # 5

AIN 5

6

Analog input # 6

AIN 6

7

Analog input # 7

AIN 7

8

Analog input # 8

AIN 8

GND

Analog input Ground

N/A

1

See Section 6.6 for details on setting up the DAS.

3.3.1.3. Connecting Analog Outputs

The 6200T is equipped with several analog output channels accessible through a
connector on the rear panel of the instrument. The standard configuration for these
outputs is mVDC. An optional current loop output is available for each. (Section 3.3.1.4).
When the instrument is in its default configuration, channels A1 and A2 output a signal
that is proportional to the SO

2

concentration of the sample gas. Either can be used for

connecting the analog output signal to a chart recorder or for interfacing with a
datalogger.
Output A3 is only used on the 6200T if the optional O

2

or CO

2

sensor is installed.

Channel A4 is special. It can be set by the user (refer to Section 5.9.9) to output any one of
the parameters accessible through the <TST TST> buttons of the unit’s sample display.
To access these signals attach a strip chart recorder and/or data-logger to the appropriate
analog output connections on the rear panel of the analyzer.

ANALOG OUT

A1 A2 A3 A4
+ - + - + - + -

Figure 3-7: Analog Output Connector

Table 3-6: Analog Output Pin Assignments

PIN

ANALOG OUTPUT

VOLTAGE SIGNAL

CURRENT SIGNAL

1

A1

V Out

I Out +

2

Ground

I Out -

3

A2

V Out

I Out +

4

Ground

I Out -

5

A3

(Only used if an

optional O

2

or CO

2

sensor is installed)

V Out

I Out +

6

Ground

I Out -

7

A4

V Out

I Out +

8

Ground

I Out -

Advertising