Remote operation using the external serial i/o, Terminal operating modes, 6200t total sulfur analyzer remote operation – Teledyne 6200T - Sulfides Analyzer User Manual
Page 179
6200T Total Sulfur Analyzer
Remote Operation
Teledyne Analytical Instruments
179
CONTROL IN
A B C D E F U
+
SP
A
N
ZER
O
Figure 8-2: Control Inputs with Local 5 V Power Supply
-
CONTROL IN
+
5 VDC Power
Supply
SPA
N
ZE
R
O
A B C D E F U
+
Figure 8-3: Control Inputs with External 5 V Power Supply
8.2. Remote Operation Using the External Serial I/O
8.2.1. Terminal Operating Modes
The 6200T can be remotely configured, calibrated or queried for stored data through the
serial ports. As terminals and computers use different communication schemes, the
analyzer supports two communication modes specifically designed to interface with
these two types of devices.
Computer mode is used when the analyzer is connected to a computer with a dedicated
interface program such as AICOM. More information regarding AICOM can be found
in later in this section or on the Teledyne’s website at http://www.teledyne-
api.com/software/apicom/.
Interactive mode is used with a terminal emulation programs such as HyperTerminal or
a “dumb” computer terminal. The commands that are used to operate the analyzer in this
mode are listed in Table 8-3 and in Appendix A-6.