6200t total sulfur analyzer calibration procedures, Teledyne analytical instruments 203, Figure 9-12: auto cal – user defined sequence – Teledyne 6200T - Sulfides Analyzer User Manual
Page 203
6200T Total Sulfur Analyzer
Calibration Procedures
Teledyne Analytical Instruments
203
To program the sample sequence shown in Table 9-4:
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL CALZ CZLS
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG
ACAL
DAS RNGE PASS CLK MORE EXIT
SETUP X.X
SEQ 1) DISABLED
NEXT
MODE
EXIT
SETUP X.X
SEQ 2) DISABLED
PREV NEXT
MODE
EXIT
SETUP X.X
MODE: DISABLED
NEXT
ENTR EXIT
SETUP X.X
MODE: ZERO
PREV
NEXT
ENTR EXIT
SETUP X.X
MODE: ZERO–SPAN
PREV
NEXT
ENTR
EXIT
SETUP X.X
SEQ 2) ZERO–SPAN, 1:00:00
PREV NEXT
MODE
SET
EXIT
SETUP X.X
TIMER ENABLE: ON
SET>
EDIT
EXIT
SETUP X.X
STARTING DATE: 01–JAN–02
<SET SET>
EDIT
EXIT
SETUP X.X
STARTING DATE: 01–JAN–02
0
4 SEP
0
3
ENTR
EXIT
Press number
buttons to set
day, month &
year:
Format :
DD-MON-Y
Y
SETUP X.X
STARTING DATE: 04–SEP–03
<SET
SET>
EDIT
EXIT
Default
value is
ON
SETUP C.4
STARTING DATE: 04–SEP–03
<SET
SET>
EDIT
EXIT
SETUP C.4
STARTING TIME:00:00
<SET SET>
EDIT
EXIT
Press number
buttons to set
time:
Format : HH:MM
Press
number
buttons to
set
number of
days
between
procedures
(1-367)
Press
number
buttons to set
delay time for
each iteration
of the
sequence:
HH:MM
(0 – 24:00)
SETUP C.4
DELTA DAYS: 1
<SET SET>
EDIT
EXIT
SETUP C.4
DELTA DAYS: 1
0 0 2
ENTR
EXIT
SETUP C.4
DELTA DAYS:2
<SET
SET>
EDIT
EXIT
SETUP C.4
DELTA TIME00:00
<SET SET>
EDIT
EXIT
SETUP C.4
DELTA TIME: 00:00
0 0 :3 0
ENTR
EXIT
SETUP C.4
DELTA TIEM:00:30
<SET
SET>
EDIT
EXIT
SETUP C.4
STARTING TIME:14:15
<SET
SET>
EDIT
EXIT
EXIT
returns
to the SETUP
Menu
SETUP C.4
DURATION:15.0 MINUTES
<SET SET>
EDIT
EXIT
SETUP C.4
DURATION 15.0MINUTES
3 0 .0
ENTR
EXIT
SETUP C.4
DURATION:30.0 MINUTES
<SET
SET>
EDIT
EXIT
Press
number
buttons to
set
duration for
each
iteration of
the
sequence:
Set in
Decimal
minutes
from
0.1 – 60.0
SETUP C.4
CALIBRATE: OFF
<SET SET>
EDIT
EXIT
SETUP C.4 CALIBRATE: OFF
ON
ENTR
EXIT
SETUP C.4
CALIBRATE: ON
<SET SET> EDIT
EXIT
Toggle
between
Off and
ON
SETUP C.4
SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
Figure 9-12:
AUTO CAL – User Defined Sequence