6200t total sulfur analyzer troubleshooting – Teledyne 6200T - Sulfides Analyzer User Manual

Page 241

Advertising
background image

6200T Total Sulfur Analyzer

Troubleshooting

Teledyen Analytical Instruments

241

Warning Message

Fault Condition

Possible Causes

PMT TEMP WARNING

PMT temperature is

< 2°C or > 12°C

Bad PMT thermo-electric cooler
Failed PMT TEC driver circuit
Bad PMT preamp board
Failed PMT temperature sensor
Loose wiring between PMT temperature sensor and PMT Preamp board
Malfunction of analog sensor input circuitry on motherboard

RCELL TEMP

WARNING

Sample chamber

temperature is

< 45°C or > 55°C

Bad reaction cell heater
Bad reaction cell temperature sensor
Bad relay controlling the reaction cell heater
Entire relay board is malfunctioning
I

2

C bus malfunction

REAR BOARD NOT

DET

Mother Board not detected
on power up.

Warning only appears on serial I/O COMM port(s)
Front panel display will be frozen, blank or will not respond.
Massive failure of mother board.

SAMPLE FLOW WARN

Sample flow rate is < 500
cc/min or > 1000 cc/min.

Failed sample pump
Blocked sample inlet/gas line
Dirty particulate filter
Leak downstream of critical flow orifice
Failed flow sensor/circuitry

SAMPLE PRES WARN

Sample Pressure is <10 in-
Hg or
> 35 in-Hg

1

If sample pressure is < 10 in-hg:

o

Blocked particulate filter

o

Blocked sample inlet/gas line

o

Failed pressure sensor/circuitry

If sample pressure is > 35 in-hg:

o

Blocked vent line on pressurized sample/zero/span gas supply

o

Bad pressure sensor/circuitry

SAMPLE PRES WARN

Sample Pressure is <10 in-
Hg or
> 35 in-Hg

1

If sample pressure is < 10 in-hg:

o

Blocked particulate filter

o

Blocked sample inlet/gas line

o

Failed pressure sensor/circuitry

If sample pressure is > 35 in-hg:

o

Blocked vent line on pressurized sample/zero/span gas supply

o

Bad pressure sensor/circuitry

SYSTEM RESET

The computer has
rebooted.

This message occurs at power on.
If it is confirmed that power has not been interrupted:
Failed +5 VDC power,
Fatal error caused software to restart
Loose connector/wiring

UV LAMP WARNING

The UV lamp intensity is <
600mV or > 4995 mV

UV lamp is bad
Reference detector is bad or out of adjustment.
Mother board analog sensor input circuitry has failed.
Fogged or damaged lenses/filters in UV light path
A/D converter circuitry failure
Light leak in reaction cell
Shutter solenoid stuck closed

1

Normally 29.92 in-Hg at sea level decreasing at 1 in-Hg per 1000 ft of altitude

(with no flow – pump disconnected).

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

A failure of the analyzer’s CPU, motherboard or power supplies can
result in any or ALL of the above messages.