Principles of Operation

Model 6200T Total Sulfur Analyzer

314

Teledyne Analytical Instruments

13.5.7.3. Status LEDs & Watch Dog Circuitry

Zero/Span and IZS O ptions
Sam ple/C AL Valve

IZS O ption
Perm eation T ube H eater

D ark Shutter

Sam ple C ham ber H eater

I2C W atchdog LED

Zero/Span and IZS O ptions
Zero/Span Valve

Figure 13-21: Relay Board Status LED Locations

Thirteen LEDs are located on the analyzer’s relay board to indicate the status of the
analyzer’s heating zones and valves as well as a general operating watchdog indicator.
Table 13-1 shows the state of these LEDs and their respective functionality.

Table 13-1:

Relay Board Status LED’s

LED

COLOR

FUNCTION

STATUS WHEN LIT

STATUS WHEN UNLIT

D1 RED Watchdog

circuit

Cycles On/Off every 3 seconds under control of the CPU.

D2 YELLOW

Sample chamber

(RCELL) heater

HEATING NOT

HEATING

D3, D4

YELLOW

Unused

N/A

N/A

D5

YELLOW

IZS heater (option)

HEATING

NOT HEATING

D6

YELLOW

Unused

N/A

N/A

D7

GREEN

Zero / Span Valve

Valve open to Span Gas path

Valve open to Zero Gas

(normal state)

D8

GREEN

Sample / Cal Valve

Valve open to

calibration gas path

Valve open to sample gas

inlet on rear panel

(normal state)

D9, D10

GREEN

Unused

N/A

N/A

D11

GREEN

UV Lamp Shutter

Shutter open

Shutter closed

D12-14

GREEN

Unused

N/A

N/A