Communications Setup

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

156

To assign or reset the status flag bit assignments, access the COMMUNICATIONS
Menu (refer to Figure 5-13), then press:

SETUP X.

HESSEN STATUS FLAGS

<SET

SET>

EDIT

EXIT

SETUP X.X

COMMUNICATIONS MENU

ID INET

HESN

COM1

COM2

EXIT

SETUP X.

PMT DET WARNING: 0002

PREV NEXT

EDIT

PRNT

EXIT

Repeat pressing SET> until …

SETUP X.

SYSTEM RESET: 0000

PREV NEXT

EDIT

PRNT

EXIT

Press <CH and

CH>

to move the

[ ]

cursor left

and right along

the bit string.

Repeat pressing NEXT or PREV until the desired

message flag is displayed. Refer to Table 7-15.

For Example …

ENTR accepts the new

settings

EXIT ignores the new

settings

SETUP X.

SYSTEM RESET: [0]000

<CH CH>

[0] ENTR EXIT

Press [?]repeatedly to cycle through the available character set: 0-9

Note: Values of A-F can also be set but are meaningless.

Figure 6-10:

COMM – Status Flag Bit Assignment

6.6.2.7. Instrument ID

Each instrument on a Hessen Protocol network must have a unique identifier (ID
number). Refer to Section 5.7.1 for information and to customize the ID of each.