Das structure, Das channels – Teledyne 6200T - Sulfides Analyzer User Manual

Page 158

Advertising
background image

DAS and AICOM

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

158

7.1. DAS Structure

The DAS is designed around the feature of a “record”. A record is a single data point of
one parameter, stored in one (or more) data channels and generated by one of several
triggering event. The entire DAS configuration is stored in a script, which can be edited
from the front panel or downloaded, edited and uploaded to the instrument in form of a
string of plain-text lines through the communication ports.
DAS data are defined by the PARAMETER type and are stored through different
triggering EVENTS in data CHANNELS, which relate triggering events to data
parameters and define certain operational functions related to the recording and
reporting of the data.

7.1.1. DAS Channels

The key to the flexibility of the DAS is its ability to store a large number of
combinations of triggering events and data parameters in the form of data channels.
Users may create up to 20 data channels and each channel can contain one or more
parameters. For each channel one triggering event is selected and up to 50 data
parameters, which can be the same or different between channels. Each data channel has
several properties that define the structure of the channel and allow the user to make
operational decisions regarding the channel (refer to Table 7-2).

Table 7-2: DAS Data Channel Properties

Property Description

Default

Setting

Range

NAME

The name of the data channel.

“NONE”

Up to 6 letters and digits
(more with AICOM, but only
the first six are displayed on
the front panel).

TRIGGERING

EVENT

The event that triggers the data channel to measure
and store its data parameters. Refer to APPENDIX
A-5 for a list of available triggering events.

ATIMER

Any allowed event.

NUMBER AND

LIST OF

PARAMETERS

A User-configurable list of data types to be
recorded in any given channel. Refer to APPENDIX
A-5 for a list of available parameters

1 - PMTDET Any available concentration,

temperature, pneumatic or
diagnostic parameter.

REPORT PERIOD The amount of time between each channel data

point.

000:01:00 000:00:01

to

366:23:59
(Days:Hours:Minutes)

NUMBER OF

RECORDS

The number of reports that will be stored in the data
file. Once the specified limit has been exceeded,
the oldest data are over-written to make space for
new data.

100

1 to 1 million, limited by
available storage space.

RS-232 REPORT Enables the analyzer to automatically report

channel values to the RS-232 ports.

OFF OFF

or

ON

CHANNEL
ENABLED

Enables or disables the channel. Provides a
convenient means to temporarily disable a data
channel.

ON OFF

or

ON

CAL HOLD OFF

Disables sampling of data parameters while
instrument is in calibration mode.
Note that - when enabled here - there is also a
length of the DAS HOLD OFF after calibration
mode, which is set in the VARS menu (refer to
Section 7.2.11).

OFF OFF

or

ON