Teledyne 6200T - Sulfides Analyzer User Manual

Page 276

Advertising
background image

Troubleshooting

Model 6200T Total Sulfur Analyzer

276

Teledyne Analytical Instruments

15. Set the front panel display to show STABIL (refer to Section 4.1.1).

16. Feed span gas into the analyzer.

17. Wait until the STABIL value is below 0.5 ppb.

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

Use a span gas equal to 80% of the reporting range.
Example: for a reporting range of 500 ppb, use a span gas of 400 ppb.

18. Scroll to the OFFSET function and record the value.

19. Scroll to the NORM PMT value.

ATTENTION

COULD DAMAGE INSTRUMENT AND VOID WARRANTY

Do not overload the PMT by accidentally setting both adjustment
switches to their maximum setting. This can cause permanent damage
to the PMT.

20. Determine the target NORM PMT value according to the following formulas.

 If the reporting range is set for  2,000 ppb (the instrument will be using the

2,000 ppb physical range):

Target NORM PMT =

(2 x span gas concentration) + OFFSET

 If the reporting range is set for  2,001 ppb(the instrument will be using the

20,000 ppb physical range):

Target NORM PMT =

(0.181 x span gas concentration) + OFFSET

EXAMPLE

: If the OFFSET IS 33 mV, the Reporting Range is 500 ppb, the

span gas should be 400 ppb and the calculation would be:

Target NORM PMT =

(2 x 400) + 33 mV

Target NORM PMT =

833 mV

21. Set the HVPS coarse adjustment switch to the lowest setting that will give you more

than the target NORM PMT signal from Step 16.

 The coarse adjustment typically increments the NORM PMT signal in 100-

300 mV steps.

22. Adjust the HVPS fine adjustment such that the NORM PMT value is at or just above

the target NORM PMT signal from Step 16.

23. Continue adjusting the both the coarse and fine switches until norm PMT is as close

to (but not below) the target NORM PMT signal from Step 16.

24. Adjust gain adjustment potentiometer until the NORM PMT value is ±10 mV of the

target level from Step 16.

25. Perform span and zero-point calibrations (refer to Section 9) to normalize the sensor

response to its new PMT sensitivity.

26. Review the slope and offset values, and compare them to the values in Table 9-5.

Advertising