6200t total sulfur analyzer calibration procedures, Step three : perform the zero/span calibration – Teledyne 6200T - Sulfides Analyzer User Manual

Page 191

Advertising
background image

6200T Total Sulfur Analyzer

Calibration Procedures

Teledyne Analytical Instruments

191

This sequence causes the

analyzer to prompt for the

expected SO

2

span

concentration.

M-P CAL

SO2 SPAN CONC: 450.0 Conc

0 0 0

4 5 .0

ENTR

EXIT

M-P CAL RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > ZERO

CONC

EXIT

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST >

CAL

SETUP

The SO

2

span concentration

values automatically default

to 450.0 Conc.

To change this value to the

actual concentration of the

span gas, enter the number

by pressing the each digit

until the expected value

appears.

The span gas concentration

should always be 90% of the

selected reporting range

EXAMPLE

Reporting range = 800 ppb

Span gas conc.= 720 ppb

EXIT

ignores the new setting

and returns to the previous

display.

ENTR

accepts the new setting

and returns to the

previous display..

Figure 9-2: Setup for Manual Calibration without Z/S valve or IZS Option (Step 2)

STEP THREE: Perform the zero/span calibration: