6200T Total Sulfur Analyzer

Calibration Procedures

Teledyne Analytical Instruments

209

9.10.2. CO

2

Sensor Calibration

Calibration begins with connecting the zero and span gases, then setting the
concentration values.

9.10.2.1. CO

2

Calibration Setup

Bottled gases are connected as follows:

Source of

SAMPLE GAS

(Remove during

calibration)

SAMPLE

EXHAUST

PUMP

3-way

Valve

Manual

Control Valve

VENT

here if input

is pressurized

Chassis

Figure 9-17:

CO

2

Sensor Calibration Set Up

CO

2

SENSOR ZERO GAS: Teledyne Analytical Instruments recommends using pure

N

2

when calibration the zero point of your CO

2

sensor option.

CO

2

SENSOR SPAN GAS: Teledyne Analytical Instruments recommends using 16%

CO

2

in N

2

when calibration the span point of your CO

2

sensor is 20%.