Teledyne 6200T - Sulfides Analyzer User Manual

Page 196

Advertising
background image

Calibration Procedures

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

196

Step Three: Perform the calibration or calibration check according to the following
flow chart:

Wait until STABIL

falls below 0.05

ppb. This may

take several

minutes.

ZERO CAL M

STABIL=XXX.X PPB

SO2 =XXX.X

< TST TST >

ZERO

CONC

EXIT

returns to the

main SAMPLE

display

Pressing ENTR changes the calibration of the instrument.

If either the ZERO or

SPAN

button fails to

appear, see Chapter 11

for troubleshooting tips.

Analyzer enters ZERO

CAL

mode.

EXIT

returns the unit to

SAMPLE

mode without

changing the calibration

values.

ZERO CAL M STABIL=XXX.X PPB

SO2 =XXX.X

< TST TST > ZERO CONC

EXIT

ZERO CAL M STABIL=XXX.X PPB SO2=X.XX

X

< TST TST > CAL

CALZ

CALS

SETUP

SPAN CAL M

STABIL=XXX.X PPB

SO2 =XXX.X

< TST TST >

SPAN

CONC

EXIT

EXIT

returns to the

SAMPLE

mode without

changing the calibration

values.

ACTION:

Allow zero gas to enter the sample port at the

rear of the instrument.

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST

TST >

CAL

CALZ

CALS

SETUP

SAMPLE

STABIL=XXX.X PPB

SO2 =XXX.X

< TST TST > CAL

CALZ

CALS

SETUP

Scroll the display to the

STABIL

test function. This

function calculates the stability

of the SO

2

measurements.

The value of STABIL

may jump

significantly. Wait

until it falls below 0.5

ppb. This may take

several minutes.

Pressing ENTR changes the calibration of the instrument.

Analyzer enters SPAN

CAL

Mode.

SPAN CAL M STABIL=XXX.X PPB

SO2 =XXX.X

< TST TST >

ENTR

CONC

EXIT

SPAN CAL M STABIL=XXX.X PPB

SO2 =XXX.X

< TST TST >

SPAN

CONC

EXIT

ZERO CAL M STABIL=XXX.X PPB

SO2 =XXX.X

< TST TST >

ENTR

CONC

EXIT

Figure 9-7: Setup for Manual Calibration with Z/S Valve Option Installed (Step 3)